国产亚洲首页精品_亚洲中文字幕久久久一区_丰满人妻av中文字幕_91丝袜人妻一区二区三区_日本国产欧美一二区

Hotline: 0755-29078051
You are here:Home >> Products >> Spectrometers >> Fiber Spectrome...

Products

PRODUCT CATAGORIES

Contact Us

Talk with me

OFFICE HOUR

 Monday - Friday : 8:30-17:30
 

CONTACT US

Hotline: 0755-29078051
E-mail:sales@glt910.com
Address: Room 201, No.1301-8 Guanguang Road, Longhua District,Shenzhen City, Guangdong Province,518110,China

Fiber Spectrometers for Industrial Control with High Sensitivities GLA639

VIS fiber spectrometer with High Sensitivities

ModelGLA639
BrandGlit
PlaceShenzhen

  • Czerny-Turner optical structure with a grating, high end light detector and SMA905 fiber connector.

    Connects to a computer through USB, RS232 or other connector for industrial control, controlled and powered by the computer.

    Unique software features, automatic configuration of spectrometer, automatic reading of calibration coefficients, spectral acquisition and measurement.

    Spectral measurements including radiation, transmittance, absorbance and reflectivity.  

    Two output data formats, conventional Excel two-column format and square format, the latter can be used to check spectral data easily which is a unique design developed by us.

    The selected spectral data on the spectral curve are marked using different colors, including the cursors and the area boundary lines, allowing users to easily find data in a large number of spectral data.

    Cursors are used to find the wavelengths. The spectral width is found by moving the two boundary lines.

    Software logging allows the user to record experiment’s content such as the material analyzed, the operator and time etc.

    Color bars help users understand the corresponding wavelengths of the spectrum.

    Interfaces and data formats for industrial control, such as Linux, ARM and PLC etc.

    GLA639 fiber spectrometer includes UV bands spectrometer, VIS band spectrometer and NIR band spectrometer.  


  • Excellent Consistency of mass production for industrial applications  

    Excellent consistency of thermal stability and vibration resistance

    Compact size and sleek appearance.


  • Items Specifications Remarks
    Fiber Connector SMA905
    Entrance Slit 10, 25, 50, 100µm
    Wavelength Range 350-1000nm Depends on the grating used
    Spectral Resolution FWHM 1.1nm@576nm@25µm slit Depends on wavelengths and slits used
    Order Sorting Filter To eliminate higher order spectrum
    Stray Light 0.06%@436nm Depends on wavelengths
    Light Detector High end detector
    Pixel Number 2048
    Pixel Size 14μmx200μm
    Detector Collecting Lens Yes
    SNR 800:01:00 Room temperature
    ADC Resolution 16 bits
    Integration Time 1 ms-15min Depends on detector used
    Data Output Count vs. wavelength or pixel
    Power Supply DC5V/380mA
    Trigger Mode Trigger in or out with TTL
    Communication USB or RS232
    Operation Software SPEC-GLA600
    Application Programming Interface glaDevSys For users to develop software
    Dimensions 100mmx65mmx36mm
    Weight 285g
    Mounting 4xM3 screw holes on the bottom
    Operating Temperature -10-50
    Storage Temperature -20-70


  • Environmental inspection: inspection of water quality and air pollution etc.

    Chemical industry  

    Optical coating inspection

    Glass and transparent materials inspection

    Measuring light sources

    R&D in science and technology

    Customization


  • Please submit your basic information, we will reply you as soon as possible!

    *

    *

    *

    *

Related products

Related applications